Troubleshooting the DEXIS Titanium sensor

If the sensor is not working, you have a couple of options you can use to troubleshoot the issue.

App

Verify that the device app is running by looking in the system tray. Ensure that the app's icon is green.

Device Manager

To ensure that the DEXIS Titanium sensor is detected by your computer, open Device Manager, and then verify that the device appears.

Sensor Hardware Errors

To view any hardware error codes, navigate to C:\ProgramData\DEXIS Imaging\Polaris\Log.

Error Code

What to Do

7 – DETECTOR_CURRENT_TOO_HIGH

Bypass any extension cables. If possible, plug sensor into multiple ports on front and back of computer. If you still receive this error, refer to DEXIS for a hardware issue.

9 – DARK_IMAGE_MEAN_TOO_HIGH

Indicates an issue with the sensor; refer to DEXIS for a hardware issue.

10 – DARK_IMAGE_MEAN_TOO_LOW

Indicates an issue with the sensor; refer to DEXIS for a hardware issue.

70 – NO_SENSOR_ATTACHED

Bypass any extension cables. If possible, plug sensor into multiple ports on front and back of computer. If  you still receive this error, refer to DEXIS for a hardware issue.